LCR-8000 Series User Manual
34
Setting DC bias
LCR-8200 Series offers DC BIAS ±12V. When input is above the
instrument voltage range will be displayed “Out of range!”.
Steps
1. Press the Measure button to enter
[MEAS DISPLAY] page.
2. Use arrow keys to move cursor and
select BIAS item on the [MEAS
DISPLAY] page.
3. Select option key on the right of the LCD
screen to adjust the frequency value by rough
or minor adjusting method or use the numeric
keypad to enter the DC Bias voltage.
4. Press the DC Bias button to output the
voltage and the DC Bias button will
illuminate. When the DC Bias is
output, the asterisk is displayed next
to the BIAS V unit.