LCR-8000 Series User Manual
38
Setting trigger delay timer
LCR-8200 Series can set the delay time before each test by setting
trigger delay timer.
The delay time range is 0ms~5000ms.
Steps
1. Press the Setup button to enter
[MEASURE MODE SETUP] page.
2. Use arrow keys to move cursor and
select TRIGGER DELAY on the
[MEASURE MODE SETUP] page.
3. Use key pad to input delay timer value ,unit is
ms.