LIST Subsystem
:LIST:STEP{1|2|3|4|…|14|15}
:LIST:STEP?
Function It sets or queries the step to edit.
Description
Set parameter
step
number1~15
Set syntax
:LIST:STEP 1
Query syntax
:LIST:STEP?
Return
data 1(Format is in <NR1>)
:LIST:PARAmeter
{OFF|RDC|LS|LP|CS|CP|Q|D|RS|RP|Z|DEG|RAD|R|X|Y|G|B}
:LIST:PARAmeter?
Function It sets or queries the measurement parameter of list
mode.
Description
Set syntax
:LIST:PARAmeter
Z
Query syntax
:LIST:PARAmeter?
Return
data Z(Format is in <disc>)
:LIST:FREQuency <frequencyNR3/disc>
:LIST:FREQuency?
Function It sets or queries the measurement frequency of list mode.
Description
Set parameter
frequency10.0~30000000.0(depending on the
model)
MAXimum/MINimum
Set syntax
:LIST:FREQuency 1000
:LIST:FREQuency 1K