208
{
0.0002~0.02
|MAXimum|MINimum}.
The stop value of
current AC
RO 25Ω is
{
0.0002~0.04
|MAXimum|MINimum}.
Set syntax
:SWEep:STARt 0.5
:SWEep:STARt 500M
:SWEep:STARt 5E-3
:SWEep:STARt MAXimum
:SWEep:STARt MINimum
Query syntax
:SWEep:STARt?
Return
data +5.000000E-01(Format is in <NR3>)
:
SWEep:FREQuency
<frequencyNR3/disc>
:SWEep:FREQuency?
Function It sets or queries the FREQ, when the TYPE setup is VAC
or IAC.
Description
Set parameter The
value of frequency is
{10.0~30000000.0|MAXimum|MINimum}.
Set syntax
:SWEep:FREQuency 1000
:SWEep:FREQuency 1K
:SWEep:FREQuency 1KHZ
:SWEep:FREQuency 1E3
:SWEep:FREQuency MAXimum
:SWEep:FREQuencyMINimum
Query syntax
:SWEep:FREQuency?
Return
data 1.000000E+03(Format is in <NR3>)
When the TYPE setup is FREQuency mode, the data returns 9.9E37.