EasyManuals Logo

GW Instek LCR Series User Manual

GW Instek LCR Series
250 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #181 background imageLoading...
Page #181 background image
COMMAND OVERVIEW
181
MEASure Subsystem
:MEASure:PARAmeter
{OFF|RDC|LS|LP|CS|CP|Q|D|RS|RP|Z|DEG|RAD|R|X|Y|G|B|E|U}
:MEASure:PARAmeter?
Function It sets or queries the measurement parameter (max.4
items) at present.
Description
Set parameter OFF, RDC(DC Resistance), Ls(Series Inductance),
Lp(Parallel Inductance), Cs (Series Capacitance), Cp (Parallel
Capacitance), Q (Quality Factor), D (Dissipation Factor), Rs (Series
Resistance) Rp(Parallel Resistance), Z(Impedance), θd(Angle),
θr(Diameter), R(Resistance), X(Reactance), Y (Admittance), G
(Conductivity), B (Susceptance).
Set syntax
:MEASure:PARAmeter RDC,Z,DEG,OFF
RDC is the first measure parameter, the second is Z, the third is
θdeg
and the forth is OFF.
Query syntax
:MEASure:PARAmeter?
Return data RDC,Z,DEG,OFF
:
MEASure
:
FREQuency
<
frequency NR3/disc>
:MEASure:FREQuency?
Function It sets or queries the FREQUENCY.
Description
Set parameter The
setup value of frequency is
{10.0~30000000.0(depending on the model)|
|MAXimum|MINimum}.
Set syntax
:MEASure:FREQuency 1000
:MEASure:FREQuency1K
:MEASure:FREQuency 1KHZ
:MEASure:FREQuency 1E3
:MEASure:FREQuency MAXimum

Table of Contents

Other manuals for GW Instek LCR Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek LCR Series and is the answer not in the manual?

GW Instek LCR Series Specifications

General IconGeneral
BrandGW Instek
ModelLCR Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals