COMMAND OVERVIEW
189
:MEASure:COMParator:UPPER <upper limit NR3>
:MEASure:COMParator:UPPER?
Function It sets or queries the UPPER.
Description
Set syntax
:MEASure:COMParator:UPPER 1
:MEASure:COMParator:UPPER1000M
:MEASure:COMParator:UPPER 1E+00
Query syntax
:MEASure:COMParator:UPPER?
Return
data 1.000000E+00 (Format is in <NR3>)
:MEASure:COMParator:LOWER <LOWER limit NR3>
Function It sets or queries the LOWER.
Description
Set syntax
:MEASure:COMParator:
LOWER
-1
:MEASure:COMParator:
LOWER
-1000M
:MEASure:COMParator:
LOWER
-1E+00
Query syntax
:MEASure:COMParator:
LOWER
?
Return
data -1.000000E+00 (Format is in <NR3>)
:MEASure:COMParator:DISPlay
<ABSolute|DEViation|PERCent|0|1|2>
:MEASure:COMParator:DISPlay?
Function It sets or queries the DISP MODE.
Description
Set parameter
ABSolute,0|DEViation,1|PERCent,2
Set syntax
:MEASure:COMParator:MODE DEViation
Query syntax
:MEASure:COMParator:MODE?
Return
data DEV(Format is in <disc>)