LCR-8000 Series User Manual
202
:LIST:OIMPedance {100|25}
:LIST:OIMPedance?
Function It sets or queries the RO of list mode.
Description
Query syntax
:LIST:OIMPedance?
Return data
100|25 (Format is in <NR1>)
:LIST:ALC {OFF|ON|0|1}
Function It sets or queries the ALC of list mode.
Description
Query syntax :LIST:ALC?
Return
data When the setup is OFF, it returns 0.
When the setup is ON, it returns 1.
:LIST:BEEPer {OFF|PASS|OK|FAIL|NG|0|1|2]
:LIST:BEEPer?
Function It sets or queries the BEEP WHEN of list mode.
Description
Set parameter
OFF, PASS|OK, FAIL|NG
Set syntax
:LIST:BEEPerPASS
:LIST:BEEPer NG
:LIST:BEEPer OFF
Query syntax
:LIST:BEEPer?
Return
data
PASS |FAIL |OFF (Format is in <disc>)
:LIST:RANGe {AUTO|HOLD|0|1}
:LIST:RANGe?
Function It sets or queries the RANGE of list mode.
Description
Set parameter If selecting <ON>, the measuring range is hold at the
range, which is used by the first time measuring. If selecting <OFF>,