LCR-8000 Series User Manual
138
Measure load
The short correction compensates for any residual impedances,
such as the impedance of the cables and the DUT connection
points.
1. Press the Measure button to enter
[MEAS DISPLAY] page, and then
press the CORR. function button on
the right side of the LCD to enter the
[CORRECTION] page.
2. Use arrow keys to move cursor and
select LOAD FUNCTION item on the
[CORRECTION] page.
3. Connect parts and test fixtures.
4. Use option key on the right of the LCD screen
to select LOAD item for LOAD data metering.
5. The LOAD item displays the measured value.