EasyManua.ls Logo

GW Instek LCR Series - Sweep:start <Startnr3;Disc; Sweep:start; Sweep:stop <Stopnr3;Disc; Sweep:stop

GW Instek LCR Series
250 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
COMMAND OVERVIEW
207
:SWEep:STARt <startNR3/disc>
:SWEep:STARt?
Function It sets or queries the START.
Description
Set parameter The start value of frequency is
{10.0~30000000.0|MAXimum|MINimum}.
The start value of
voltage AC
RO 100Ω is
{0.01~2|MAXimum|MINimum}.
The start value of
voltage AC
RO 25Ω is
{0.01~1|MAXimum|MINimum}.
The start value of
current AC
RO 100Ω is
{
0.0002~0.02
|MAXimum|MINimum}.
The start value of
current AC
RO 25Ω is
{
0.0002~0.04
|MAXimum|MINimum}.
Set syntax
:SWEep:STARt 0.5
:SWEep:STARt 500M
:SWEep:STARt 5E-3
:SWEep:STARt MAXimum
:SWEep:STARt MINimum
Query syntax
:SWEep:STARt?
Return
data +5.000000E-01(Format is in <NR3>)
:SWEep:STOP <stopNR3/disc>
:SWEep:STOP?
Function It sets or queries the STOP.
Description
Set parameter The stop value of frequency is
{10.0~30000000.0|MAXimum|MINimum}.
The stop value of
voltage AC
RO 100Ω is
{0.01~2|MAXimum|MINimum}.
The stop value of
voltage AC
RO 25Ω is
{0.01~1|MAXimum|MINimum}.
The stop value of
current AC
RO 100Ω is

Table of Contents