COMMAND OVERVIEW
207
:SWEep:STARt <startNR3/disc>
:SWEep:STARt?
Function It sets or queries the START.
Description
Set parameter The start value of frequency is
{10.0~30000000.0|MAXimum|MINimum}.
The start value of
voltage AC
RO 100Ω is
{0.01~2|MAXimum|MINimum}.
The start value of
voltage AC
RO 25Ω is
{0.01~1|MAXimum|MINimum}.
The start value of
current AC
RO 100Ω is
{
0.0002~0.02
|MAXimum|MINimum}.
The start value of
current AC
RO 25Ω is
{
0.0002~0.04
|MAXimum|MINimum}.
Set syntax
:SWEep:STARt 0.5
:SWEep:STARt 500M
:SWEep:STARt 5E-3
:SWEep:STARt MAXimum
:SWEep:STARt MINimum
Query syntax
:SWEep:STARt?
Return
data +5.000000E-01(Format is in <NR3>)
:SWEep:STOP <stopNR3/disc>
:SWEep:STOP?
Function It sets or queries the STOP.
Description
Set parameter The stop value of frequency is
{10.0~30000000.0|MAXimum|MINimum}.
The stop value of
voltage AC
RO 100Ω is
{0.01~2|MAXimum|MINimum}.
The stop value of
voltage AC
RO 25Ω is
{0.01~1|MAXimum|MINimum}.
The stop value of
current AC
RO 100Ω is

Table of Contents

Other manuals for GW Instek LCR Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek LCR Series and is the answer not in the manual?

GW Instek LCR Series Specifications

General IconGeneral
BrandGW Instek
ModelLCR Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals