RO (output impedance) is setting 25Ω, the setup value of AC
voltage is {0.01~1|MAXimum|MINimum}.
Set syntax
:MEASure:VOLTage:AC 1
:MEASure:VOLTage:AC 1M
:MEASure:VOLTage:AC 1MV
:MEASure:VOLTage:AC 1E-3
:MEASure:VOLTage:AC MAXimum
:MEASure:VOLTage:AC MINimum
Query syntax
:MEASure:VOLTage:AC?
Return data 1.000000E+00 (Format is in <NR3>)
When the Level setup is current mode, the data returns 9.9E37.
:MEASure:CURRent:AC <current NR3/disc>
:MEASure:CURRent:AC?
Function It sets or queries the AC measurement current (LEVEL).
Description
Set parameter When RO(output impedance) is setting 100Ω, the
setup value of AC current is {0.0002~0.02|MAXimum|MINimum}.
When RO (output impedance) is setting 25Ω, the setup value of AC
current is {0.0002~0.04|MAXimum|MINimum}.
Set syntax
:MEASure:CURRent:AC 0.01
:MEASure:CURRent:AC 10M
:MEASure:CURRent:AC 10MA
:MEASure:CURRent:AC 1E-2
:MEASure:CURRent:AC MAXimum
:MEASure:CURRent:ACMINimum
Query syntax
:MEASure:CURRent:AC?
Return data 1.000000E-02 (Format is in <NR3>)
When the Level setup is voltage mode, the data returns 9.9E37.

Table of Contents

Other manuals for GW Instek LCR Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek LCR Series and is the answer not in the manual?

GW Instek LCR Series Specifications

General IconGeneral
BrandGW Instek
ModelLCR Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals