MEASURE (METER MODE)
39
Setting AC/DC delay timer
LCR-8200 Series can set the AC/DC delay time when Rdc
parameter enable .
The inductance and Rdc values can be measured and displayed
simultaneously.
When the inductance is measured by Rdc, a current flows through
the generated magnetic field. When the DC signal ends, the
inductance will generate a back electromotive force. If the AC
signal is subsequently sent out for measurement, the L value may
cause an error. To avoid this problem, you can set the AC/DC
delay timer to reduce the effect of back EMF on the measurement.
This delay timer will be executed both AC to DC and DC to AC.
The delay time range is 0ms~5000ms.
1. Press the Setup button to enter
[MEASURE MODE SETUP] page.
2. Use arrow keys to move cursor and
select AC/DC DELAY on the
[MEASURE MODE SETUP] page.
3. Use key pad to input delay timer value ,unit is
ms.