:CORRection:SHORt
:CORRection:SHORt?
Function It sets or queries to do short correction.
Description
Query syntax
:CORRection:SHORt?
Return
data When correction fail, it returns 0.
When correction passes, it returns 1.
:CORRection:OPEN:STATe {OFF|ON|0|1}
:CORRection:OPEN:STATe?
Function It sets or queries the status of open correction.
Description
Query syntax
:CORRection:OPEN:STATe?
Return
data When the setup is OFF, it returns 0.
When the setup is ON, it returns 1.
:CORRection:SHORt:STATe {OFF|ON|0|1}
:CORRection:SHORt:STATe?
Function It sets or queries the status of short correction.
Description
Query syntax
:CORRection:SHORt:STATe?
Return
data When the setup is OFF, it returns 0.
When the setup is ON, it returns 1.
:CORRection:CABLe {0|0.5|1|2}
:CORRection:CABLe?
Function It sets or queries the status of cable length correction.
Description
Query syntax
:CORRection:CABLe?
Return
data 0|0.5|1|2

Table of Contents

Other manuals for GW Instek LCR Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek LCR Series and is the answer not in the manual?

GW Instek LCR Series Specifications

General IconGeneral
BrandGW Instek
ModelLCR Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals