Description:
Query syntax
FETC?
Return
data
Meter Mode
Read Data format under Meter Mode
<para 1 data>,<para 2 data>,<para 3 data>,<para 4
data>,<status>,<bin number>,<para 1 compare
status>,<para2 compare status>,<para 3 compare
status>,<para 4 compare status>
para 1-4 data
During value measuring, not all of four values will be
displayed. The value will be displayed when the parameters
are available. For example, if only two parameters are
enabled, then only two values will be transmitted.
status – Measuring status, and the weighted value of each
status refers to the final value.
0 – Normal status without special status and without
comparison.
1 – Measuring schedule error
2 – ALC error
4 – Other errors
8 – Reserve
16 – All parameters OK
32 – Some parameters NG
bin number – Categorization result, and such value will not