LCR-8000 Series User Manual  
66 
Setting level/freq 
When performing a frequency sweep test, you need to set the test 
voltage/current (Vac/Iac). If you are performing a voltage/current 
scan test, you need to set the test frequency. The voltage range is 
10mV-2Vrms and current range is 0.1mA-20mArms. 2Vrms can 
only be used at <= 1MHz. 
 
1.  Press the Measure button to enter 
[MEAS DISPLAY] page, and then press 
the SWEEP function button on the 
right side of the LCD to enter the 
[SWEEP DISPLAY] page. 
2.  Use arrow keys to move the cursor and 
select LEVEL/FREQ item on the 
[SWEEP DISPLAY] page. 
 
 
3.   Use the numeric keypad to enter the test 
Vac/Iac (frequency) and unit.