LCR-8000 Series User Manual
198
:LIST:DELayMINimum
Query syntax
:LIST:DELay?
Return data 0.500 (Format is in <NR2>)
:LIST:COMParator:MODE <ABSolute|DEViation|PERCent|0|1|2>
Function It sets or queries the COMP MODE of list mode.
Description
Set parameter
ABSolute,0|DEViation,1|PERCent,2
Set syntax
:MEASure:COMParator:MODE PERCent
Query syntax
:LIST:COMParator:MODE?
Return data PERC (Format is in <disc>)
:LIST:COMParator:NOMinal <nominal value NR3 >
:LIST:COMParator:NOMinal?
Function It sets or queries the NOMINAL of list mode.
Description
Set syntax
:LIST:COMParator:NOMinal 1000
:LIST:COMParator:NOMinal 1K
:LIST:COMParator:NOMinal 1E+03
Query syntax
:LIST:COMParator:NOMinal?
Return
data 1.000000E+03 (Format is in <NR3>)
:LIST:COMParator:UPPER <upper limit NR3>
:LIST:COMParator:UPPER?
Function It sets or queries the UPPER of list mode.
Description
Set syntax
:LIST:COMParator:UPPER 1
:LIST:COMParator:UPPER1000M
:LIST:COMParator:UPPER 1E+00
Query syntax
:LIST:COMParator:UPPER?
Return data 1.000000E+00 (Format is in <NR3>)

Table of Contents

Other manuals for GW Instek LCR Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GW Instek LCR Series and is the answer not in the manual?

GW Instek LCR Series Specifications

General IconGeneral
BrandGW Instek
ModelLCR Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals