GW Instek LCR Series - Measure:range:ac {1|2|3|4|AUTO|HOLD; Measure:oimpedance {100|25; Measure:oimpedance; Measure:comparator:parameter {1|2|3|4

GW Instek LCR Series
250 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
COMMAND OVERVIEW
187
:MEASure:RANGe:AC {1|2|3|4|AUTO|HOLD}
Function It sets or queries the AC RANGE.
Description
Set parameter {1|2|3|4|
AUTO
|
HOLD
}
Set syntax
:MEASure:RANGe:AC 3
:MEASure:RANGe:AC AUTO
Query syntax
:MEASure:RANGe:AC
Return data 3 (Format is in <NR1>)
:MEASure:OIMPedance {100|25}
:MEASure:OIMPedance?
Function It sets or queries the RO.
Description
Query syntax
:MEASure:OIMPedance?
Return
data 100|25 (Format is in <NR1>)
:MEASure:COMParator:PARAmeter {1|2|3|4}
:MEASure:COMParator:PARAmeter?
Function It sets or queries which parameters number is to be
compared.
Description
Set syntax
:MEASure:COMParator:PARAmeter 3 (After this
command, every compared setup is at this parameters number.)
Query syntax
:MEASure:COMParator:PARAmeter?
Return
data 3 (Format is in <NR1>)