The open correction is quite sensitive to noise - 
both noise originating externally and induced 
noise. Therefore, if open correction fails, you 
should check the following points before starting 
the correction process again: 
  Check that the test fixture or the test cables are 
properly connected. 
  Check that not any test piece is connected to 
the test fixture or the test cables.  
  During the correction process, be sure not to 
disturb the fixture and the test cables or to 
move your hand near them. 
  If the H/L test pole of the test fixture can be 
adjusted, adjust the distance between the test 
poles to the distance when measuring.