7.4 Taking Measurements
────────────────────────────────────────────────────
.4.1 Clearing Measurement Data and Setting the Range
NOTE
NOTE
.4 Taking Measurements
(1) Clearing measurement data
Press the DATA RESET key to clear all of the measurement data.
Detection of the integrated value and minimum/maximum values does not
occur until the demand measurement operation is started.
(2) Setting the range
The measurement range is fixed while the demand measurement operation is
in progress.
Set a suitable voltage/current range (fixed range) before starting integrated
measurement.
・When "Auto" is set for the measurement range:
The measurement range is fixed at the moment that the START/STOP key is
pressed. Note that correct measurement is impossible if there is any input
that is outside of the fixed range that was set. Once the demand
measurement operation is completed, the measurement range setting does not
return to "Auto."
・If a minimum or maximum value is generated while the power meter is in
standby before the start of demand measurement, that value is not retained.