170
────────────────────────────────────────────────────
11.4 Command Reference
────────────────────────────────────────────────────
Display averaging time setting
Synta
x
AVERaging data
dat
a
1, 2, 3, 4, 5, 6, 7, 8, 9, 10 (numerical data in NR1 format)
Functio
n
Sets the number of measurements over which the display averaging time is
to be performed.
data can be received in NRf format, but after the decimal point 5 and above
will be rounded up while 4 and below will be rounded down.
Not
e
Setting the averaging time causes a restart.
Erro
r
If data is not in NRf format, a command error is generated.
If data is not one of the above listed values, an execution error is generated
.
In the following circumstances, a device dependent error is generated:
If this command is executed in the HOLD state .
In the event of a system error (the command is not executed).
Integration or demand measurement is in progress.
Exampl
e
Transmission
":AVER 2"
Display averaging time is set to 2.
Display averaging time query
Synta
x
AVERaging?
Functio
n
Returns the current setting of the number of measurements over which the
display averaging time is to be performed, as a numerical value (data)in
NR1 format.
Respons
e
synta
x
If headers are ON
If headers are OFF
"AVERAGING data"
"data"
Exampl
e
If headers are ON
If headers are OFF
Transmission
Response
":AVER?"
":AVERAGING 2"
":AVER?"
"2"