199
────────────────────────────────────────────────────
11.4 Command Reference
────────────────────────────────────────────────────
Integrated measurement mode FD automatic output setting
Synta
x
INTEgrate:DATAout:FD data
dat
a
ON, OFF (character data)
Functio
n
Sets Integrated measurement mode FD automatic output to ON or OFF.
Not
e
If there is no FDD unit connected, if there is no floppy disk loaded in the
drive, if an existing file name is being used, or if the disk is write-protected
,
this command itself ends normally, but the first attempt to execute
automatic output will generate an error, and automatic output will not be
executed.
Erro
r
If data is set to character data other than "ON" or "OFF", an execution
error is generated.
In the following circumstances, a device dependent error is generated:
If an attempt was made to execute this command while waiting for or
during integrated measurement.
In the event of a system error.
Exampl
e
Transmission
":INTE:DATA:FD ON"
Sets integrated mode FD automatic output to on.
Integrated measurement mode FD automatic output query
Synta
x
INTEgrate:DATAout:FD?
Functio
n
Returns the current Integrated measurement mode FD automatic output
query indataas either ON or OFF.
Respons
e
synta
x
If headers are ON
If headers are OFF
":INTEGRATE:DATAOUT:FD data"
"data"
Exampl
e
If headers are ON
If headers are OFF
Transmission
Response
":INTE:DATA:FD?"
":INTEGRATE:DATAOUT:FD ON"
":INTE:DATA:FD?"
"ON"
I
NTEgrate:DATAout:FD
I
NTEgrate:DATAout:FD?

Table of Contents