211
────────────────────────────────────────────────────
11.4 Command Reference
────────────────────────────────────────────────────
■ Demand measurement instantaneous data query
Synta
MEASure:DEMand:INSTant? data1,data2
dat
data1 0 to 255: Voltage, current instantaneous value specification
(numerical data in NR 1)
data2 0 to 255: Power instantaneous value specification
(numerical data in NR 1)
128 64 32 16 8 4 2 1
bit 7 bit 6 bit 5 bit 4 bit 3 bit 2 bit 1 bit 0
data1 Iave I3 I2 I1 Uave U3 U2 U1
data2 F PF S Q P
Functio
Returns the data for the measurement item specified by (data1) and
(data2).
Respons
synta
Headers
Data formats
U1,U2,U3,Uave,I1,I2,I3,Iave (data1)
P,Q,S,PF,F (data2)
+123.45E+67
Not
The message unit separator can be switched from a semicolon (";") to a
comma (",") with the "TRANsmit:SEParator" command.
Erro
・ In the following circumstances, a device dependent error generated:
If the measurement mode is not Demand Measurement mode
In the event of a system error.
・If a value outside of the setting range is set for either <data1> or <data2>, a
command error results.
Exampl
Transmission
":MEAS:DEM:INST? 255,0"
Inquires about "U1", "U2", "U3", "Uave", "I1" ,"I2", "I3", and "Iave" from the
measurement data.
EASure:DEMand:INSTant?