11.4 Command Reference
────────────────────────────────────────────────────
Not
・"qtime" for (data3) specifies the time at which this query was received.
・"start" for (data4) specifies the starting time of the measurement for the
minimum/maximum values.
・When "MEAS:DEMand:DEMand?" has been specified, nothing is output even
if "TLF" or "LF" is specified for (data11) and (data12) and (data13) are also
specified.
・The message unit separator can be switched from a semicolon (";") to a
comma (",") with the "TRANsmit:SEParator" command.
・If the reactive power meter method is not used, even if "WQ-" or "WQ+" is
specified for (data9), nothing is output.
・If the reactive power meter method is used, even if "WS-" or "WS+" is
specified for (data10), nothing is output.
Erro
・An execution error occurs if a nonexistent date is set.
・In the following circumstances, a device dependent error is generated:
If the measurement mode is not Demand measurement mode.
In the event of a system error.
When no items at all are specified.
Exampl
Transmission
":MEAS:DEM:DEM? 0,0,0,0,0,0,0,0,7,0,0"
Inquires about "Wh", "Wh+" and "Wh-" from the demand.