219
────────────────────────────────────────────────────
11.4 Command Reference
────────────────────────────────────────────────────
Pmax DATE,Pmax TIME,Qmax DATE,Qmax TIME,Q-max DATE,
Q-max TIME,Smax TIME,Smax DATE,PFmax DATE,PFmax TIME,
PF-max DATE,PF-max TIME,Fmax DATE,Fmax TIME,
START DATE,START TIME
data6
U1min,U2min,U3min,I1min,I2min,I3min
data7
Pmin,Qmin,Q-min,Smin,PFmin,PF-min,Fmin
data8
U1min DATE,UImin TIME,U2min DATE,U2min TIME,
U3min DATE,U3min TIME
data9
Pmin DATE,Pmin TIME,Qmin DATE,Qmin TIME,Q-min DATE,
Q-min TIME,Smin DATE,Smin TIME,PFmin DATE,PFmin TIME,
PF-min DATE,PF-minTIME,FminDATE,FminTIME
data10
WP,WP+,WP-,WQ,WQ+,WQ-,ELAPSED TIME
data11
WS,WS+,WS-
data12
Pave,Qave,Save,PFave
data13
Data formats
Date
Hours
Max, and min data,PF
,
F,PFave
Instantaneous value
data except PF,F and
Pave,Qave and Save
Integrated data
Integrated elapsed
measurement time
yy/mm/dd
hh:mm:ss
+123.4E+0(9 characters in all)
+123.45E+0(10 characters in all)
+123.456E+0(11 characters in all)
hhhh:mm:ss
Not
e
"qtime" for (data5) specifies the time at which this query was received.
"start" for (data6) specifies the starting time of the measurement for the
minimum/maximum values.
"etime" for (data11) specifies the time at which integrated elapsed
measurement time.
If the reactive power meter method is used, even if "WS-" or "WS+" is
specified for (data10), nothing is output.
If the reactive power meter method is not used, even if "WQ-" or "WQ+" is
specified for (data9), nothing is output.
The message unit separator can be switched from a semicolon (";") to a
comma (",") with the "TRANsmit:SEParator" command.
Erro
r
In the following circumstances, a device dependent error is generated:
If the measurement mode is not Integrated measurement mode.
In the event of a system error.
When no items at all are specified.
Exampl
e
Transmission
":MEAS:INTE? 255,0,0,0,0,0,0,0,0,0,0,0,0"
Inquires about "U1", "U2", "U3", "Uave", "I1", "I2", "I3" and "Iave" from the
measurement data.