Pathway Technical Reference Manual
Proprietary 183 of 190
The following safety limitation has
been violated.
Temperature: XX °C; Allowed time: X
(sec). Error Code: 8002
The temperature duration limit has been
exceeded. The PATHWAY system automatically
switches to Safe Mode in order to prevent any
damage.
Before attempting to resume normal
operation:
1. Check that the parameters of the
program used are valid.
2. Check that the Thermodes are intact and
undamaged.
3. Run the system Self-Test
If the problem persists, stop using the
PATHWAY system and contact Medoc service.
Safety cold duration limit violation.
Temperature level: # (°C), Time
entry: ## (sec). Error Code: 800E
The temperature duration limit has been
exceeded. The PATHWAY system automatically
switches to Safe Mode in order to prevent any
damage.
Before attempting to resume normal
operation:
1. Check that the parameters of the
program used are valid.
2. Check that the Thermodes are intact and
undamaged.
3. Run the system Self-Test
If the problem persists, stop using the
PATHWAY system and contact Medoc service.
Static sensor mismatch. Error Code:
8003
The system has detected a mismatch between
the Thermode sensors. PATHWAY switches to
Safe Mode in order to prevent any damage.
Before attempting to resume normal
operation:
1. Check that the Thermodes are intact and
undamaged.
2. Run the system Self-Test
3. Re-calibrate the Thermode.
If the problem persists, stop using the system
and contact Medoc service.
Static sensor mismatch. Mismatch
temperature # (°C). Error Code:
8003