283
Automated Mode Alarms 2424 Automated Mode Alarms
What to do
1. Tap NEXT to see the next screen.
2. Use a BG meter to conrm your blood glucose.
If low conrmed, consider treatment.
If high conrmed, check infusion (Pod) site and
ketones.
If your sensor glucose value is not what you
expected, you may need to calibrate or replace
your Dexcom G6 Sensor.
3. Tap NEXT aer you conrm your blood glucose.
4. Tap SWITCH TO MANUAL MODE, then stay in
Manual Mode for at least ve minutes.
While in Manual Mode, you can check your Sensor Graph to nd out whether
your insulin has been paused or has been at a maximum for a long time.
Aer at least 5 minutes of Manual Mode, you can return to Automated Mode aer
you have conrmed your sensor glucose values are accurate.
For more information about Automated Mode: Limited state, see “21.5. About
Automated Mode: Limited on page 269.