327
Appendix
Summary of Settings and Options
e options for the various Omnipod 5 Automated Insulin Delivery System
settings are:
Time format 12-hour
Time zone GMT-11:00 to GMT+13.00.
Daylight Savings Time ON or OFF. Default based on date and time zone.
Date format MM/DD/YYYY
Screen time-out 30, 60, 120 seconds. Default is 30 seconds.
PIN 4 digits from 0 to 9.
Dexcom G6 Transmitter
serial number (SN)
6 characters.
Maximum Basal Rate Select one value between 0.05-30 U/hr in 0.05 U/hr
increments. Default is 3.00 U/hr.
Basal rate Units/hr. Range: 0 U/hr to Maximum Basal Rate in
0.05 U/hr increments.
Basal Programs Maximum of 12.
Basal rate segments 24 per Basal Program.
Activity feature Range: 1 to 24 hrs
In increments of 1 hour
Temp basal %, units/hr, or OFF. Default is OFF.
Duration: 30 min to 12 hrs in 30-min increments.
Temp basal (set to %) Range: 100% decrease (0 U/hr) to 95% increase from
current basal rate in 5% increments. Cannot exceed
Maximum Basal Rate.
Temp basal (set to U/hr) Range: 0 U/hr to Maximum Basal Rate in increments of
0.05 U/hr.
Temp basal presets Maximum of 12.
Glucose Goal Range (for
blood glucose history)
Lower and upper limits: 3.9 to 11.1 mmol/L in
0.1 mmol/L increments .
BG reminder ON or OFF. Default is OFF.
Maximum of 4 active at one time.
A reminder can occur between 30 min and 4 hrs aer a
bolus is started. Set in 30-minute increments.
Target Glucose value Maximum of 8 segments; 6.1 to 8.3 mmol/L in
0.55 mmol/L increments.