Appendix
328
Appendix
Correct Above threshold Maximum of 8 segments; Target Glucose to 11.1
mmol/L in 0.1 mmol/L increments.
Minimum Glucose for
Calculations
2.8 to 3.9 mmol/L in 0.1 mmol/L increments
Default is 3.9 mmol/L.
Insulin-to-carb (IC)
ratio
Maximum of 8 segments; 1 to 150 g carb/U in
0.1 g carb/U increments.
Correction (sensitivity)
factor
Maximum of 8 segments; 0.1-22.2 mmol/L in 0.1
mmol/L increments. Default is 2.8 mmol/L.
Reverse Correction ON or OFF. Default is ON.
Duration of insulin
action
2 to 6 hours in 30-minute increments. Default is 4 hours.
Bolus size Range: 0.05-30 U in 0.05 U increments.
Extended bolus %, Units, or OFF. Default is OFF.
30 minutes to 8 hours in 30-minute increments.
Pause insulin 30 minutes to 2 hours.
Low Pod Insulin
advisory
10 to 50 units in 1-unit increments. Default is 10.0 U.
Pod expiration
notication
1 to 24 hours in 1-hour increments. Default is 4 hours.
Pod Shut-O timer OFF, or 1 to 24 hours in 1-hour increments. Default is
OFF.
History screen display Rolling 90-day period.
Language English.
Pod Specications
Size: 3.9cm wide x 5.2cm long x 1.45cm high (1.53" x 2.05" x 0.57")
Weight (without insulin): 26 grams (0.92 oz)
Operating temperature range: Pod operating environment of 5°C to 40°C (41°F
to 104°F).
Startup temperature: above 10°C (50°F)
Storage temperature range: 0°C to 30°C (32°F to 86°F)
Warm-up time (0°C to 20°C): 7 minutes
Cooldown time: No time is required for cooldown from maximum storage
temperature (30°C) to operating temperature.
Reservoir volume (deliverable): 200 units
Cannula insertion depth: 4 to 7 mm (0.16-0.28 in)
Depth of insulin infusion: ≥ 4 mm (0.16 in)