Controller Specications
Size: 143.92 mm high x 67.57 mm wide x 12.33 mm deep (5.67" x 2.66" x 0.49")
Weight: 165 grams (5.82 oz)
Screen active area: 56.16 mm wide x 120.58 mm high (2.21" x 4.75")
Operating temperature range: 5°C to 40°C (41°F to 104°F)
Storage temperature range: 0°C to 30°C (32°F to 86°F)
Operating relative humidity range: 20% to 90%, non-condensing
Storage relative humidity range: 20% to 90%, non-condensing
Operating atmospheric pressure: 700 hPA to 1060 hPA
Storage atmospheric pressure: 700 hPA to 1060 hPA
Communication distance: e Controller and Pod should be:
• At startup: Adjacent and touching, with the Pod either in or out of tray, to
ensure proper communication during priming.
• During normal operation: Within 1.5 meters (5 feet) of each other.
Depending on the location, the communication distance may handle
separations up to 15 meters (50 feet) away.
Alarm type: Audible. Output: ≥ 45 db (A) at 1 meter
IP (Ingress Protection) rating for moisture and dust: IP22 (protected from
touch by ngers and objects 12.5 millimeters or larger; not well-protected from
water - avoid liquid)
Notication type: Audible and vibratory
Battery: Rechargeable Li-ion battery, 3.8V, 2800 mAh
Battery Operational Life: Full charge covers approximately 36 hours with typical
use.
Controller Service Life: Approximately 2 years (based on 300-500 charge cycles)
with typical use
Shelf Life (Starter Kit): 18 months
Battery charger operating line voltage: 100 to 240 VAC, 50/60 Hz
Only use the Noetic approved power adapter (Insulet PN PT-000428) with
the Controller.
Dexcom Specications
For information about Dexcom operating specications, see the Dexcom G6
CGM System Instructions for Use.