Setting the Output Conditions
ZW-7000/5000
User's Manual
145
4
Settings for Function
Multi View Explore : [Bank] | [(Bank Data Name)] (double click)
→ Edit pane : [Task Settings] icon ( )
→ Task Settings window : [Hold]
Hold Hold mode Auto Peak to
Peak
Holds the difference between the maximum and minimum values of the
measurement result. Output the measurement value at each updating of the
difference value Maximum and Minimum.
Average Holds the “average of the measured value” of the sampling period. The current
measure
ment
value is output after Sampling, and then this state is held until
the next sampling ends.
Sampling Holds the measured value the moment that control enters the sampling
per
iod. The
current measurement value is output after Sampling, and then this
state is held until the next sampling is started.
Through
(default value)
The hold function is not used. The measure
d v
alue is always output.
1
Select the hold mode from [Hold mode].
Setting items for required parameters are
display
ed depending on the selected hold
mode.
Setting Triggers p.147
Setting a Trigger Delay p.151
The hold mode can also be set by the operating keys on the Sensor Controller.
Setting the Hold Mode p.225
Item Setting item Setting value Description
00
Current
measured value
Output
(Max.value - Min.value)
Current
measured value
Output
(Average)
Sampling period
Current
measured value
Output
Sampling period
Note