EasyManua.ls Logo

Omron ZW-7000 Series - Setting Threshold Value

Omron ZW-7000 Series
290 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Setting the Output Conditions
ZW-7000/5000
User's Manual
155
4
Settings for Function
Setting the Zero Reset Memory
Select whether or not to hold the measured value zero reset level even if the power is turned OFF.
As shown in the figure below, use the zero reset memory still set to “OFF” when using this function to perform a
zero reset at
each measurement.
Example: When the level difference of a measurement target is measured
When zero reset memory is set to “ON”, the zero reset is written to EEPROM (non-volatile memory) in the Sensor
Controller each time that a zero reset is performed. The EEPROM can be written a maximum of 1,000,000 times.
Pay attention to the maximum number of writing allowed when this is set to “ON.
Even if zero reset memory is disabled, the zero reset lev
el w
ill be held also when the setting is saved. In this
instance, the zero reset will also be continued after a restart.
Item Setting item Setting value Description
Sensor settings Zero reset memory OFF (default
value)
The zero reset is canceled when the power is turned OFF.
ON The zero reset level is held even if the power is turned OFF.
This step is
measured.
Sensor head
Zero reset
executed.
Direction of movement
Zero reset
executed.
Zero reset
executed.
Important

Table of Contents