Basic Operations for Displaying Measurement Results
36
ZW-7000/5000
User's Manual
Multi View Explore : [Device Group] | [(Sensor Name)] | [System] | [System Data] (double click)
→ Edit pane: [Sensor Settings] icon ( )
• The calibration data is stored inside the Sensor Controller. Therefore, if the sensor head is calibrated once, as long
as the fiber connection state does not change (if it is not disconnected from or connected to the Sensor Controller or
an extension added), this operation is not needed each time. After removing/inserting a fiber cable from/to the Sen-
sor Controller, or extending a fiber cable, be sure to calibrate the Sensor H
ead again.
• When Senso
r Head calibration fails
When sensor head calibration is performed in an inappropriate en
vironment or w
ith an extension fiber cable set
incorrectly, an error message is displayed. In such a case, refer to the p.66, recheck your environment, and try
again. If an error results even if Sensor Head calibration is done under the correct conditions, it is p
ossible that
some contaminants may have adhered to the Sensor Head part, or the fiber connector of the Sensor Controller.
Clean the fiber cable or fiber connector referring to p.59. After cleaning the fiber or fiber connector
, be sure to
calibrate the Sensor Head again.
4 Click [Execute sensor head calibration].
The sensor head calibration window appears.
5 With no object in the measuring range, or
with the tip of the sensor head shielded from
light by an object that has the characteris-
tics of diffusion reflection, click the [Execute
s
ens
or head calibration] button.
Important