ZVT GUI Reference
Channel Menu
Operating Manual 1145.1084.12 – 30 409
Off disables the chopped pulse profile mode. All other settings are left unchanged.
In remote control it is possible to skip measurement intervals (and thus accelerate the measurement,
at the expense of the number of sweep points) by increasing the Delay Increment parameter.
[SENSe<Ch>:]PULSe:GENerator2:WIDTh
[SENSe<Ch>:]PULSe:GENerator<gen_no>:CPPRofile
[SENSe<Ch>:]PULSe:GENerator<gen_no>:DINCrement
File Format for Pulse Train Definitions
The analyzer uses a simple ASCII format to export pulse train definitions. By default, the pulse train file
has the extension *.train and is stored in the directory shown in the Save Pulse Train and Recall Pulse
Train dialogs. The file starts with a preamble containing the channel, the Pulse Train Period, and the Pulse
Delay (to Sync) value. The following lines contain the entries of all editable columns of the list.
Example of a pulse train definition file
The pulse train definition:
is described by the pulse train file:
MMEMory:LOAD:PTRain
MMEMory:STORe:PTRain
Spurious Avoidance
The Spurious Avoidance submenu defines whether the analyzer measures with a local oscillator
frequency LO below or above the RF input frequency.