R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 652
CALCulate<Chn>:MARKer<Mk>:FUNCtion:TARGet <search_value>
[Deprecated command]
Defines the target value for the target search of marker no. <Mk>, which can be activated using
CALCulate<Chn>:MARKer<Mk>:FUCTion:EXECute TARGet.
Note: This command is the ZVR-compatible equivalent of CALCulate<Chn>:MARKer<Mk>:TARget.
Channel number used to identify the active trace
Marker number in the range 1 to 10
Target search value of marker no. <Mk>.
Depending on the format of the active trace (CALCulate<Chn>:FORMat). For
a dB Mag trace the range is –300 dB to +200 dB, the increment (UP. DOWN) is
0.1 dB. [dB, deg etc., depending on the trace format]
Depending on the trace format; 0 dB for a dB Mag trace
Device-specific, command or query
CALCulate<Chn>:MARKer<Mk>:MAXimum
[Deprecated command]
Selects a search mode for marker no. <Mk> and initiates a maximum search. The marker must be created
before using CALCulate<Chn>:MARKer<Mk>[:STATe] ON.
Note: This command is the ZVR-compatible equivalent of
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute MAXimum.
Channel number used to identify the active trace
Marker number in the range 1 to 10
Device-specific, no query
CALCulate<Chn>:MARKer<Mk>:MINimum
[Deprecated command]
Selects a search mode for marker no. <Mk> and initiates a minimum search. The marker must be created
before using CALCulate<Chn>:MARKer<Mk>[:STATe] ON.
Note: This command is the ZVR-compatible equivalent of
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute MINimum.
Channel number used to identify the active trace
Marker number in the range 1 to 10
Device-specific, no query