R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 657
See CALCulate<Chn>:MARKer<Mk>:BWIDth.
CALCulate<Chn>:MARKer<Mk>:SEARch[:IMMediate]
[Deprecated command]
Initiates a search according to the search function selected with
CALCulate<Chn>:MARKer<Mk>:FUNCtion[:SELect]. The marker must be created before using
CALCulate<Chn>:MARKer<Mk>[:STATe] ON.
Note: Together with CALCulate<Chn>:MARKer<Mk>:FUNCtion[:SELect] this command is the ZVR-
compatible equivalent of CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute ....
Channel number used to identify the active trace
Marker number in the range 1 to 10. For a bandfilter search (BFILter) this
numeric suffix is ignored and may be set to any value because the bandfilter
search functions always use markers M 1 to M 4.
Device-specific, no query
CALCulate<Chn>:MARKer<Mk>:SEARch:LEFT
[Deprecated command]
Selects a search mode for marker no. <Mk> and initiates a search for the next valid peak to the left. The
marker must be created before using CALCulate<Chn>:MARKer<Mk>[:STATe] ON.
Note: This command is the ZVR-compatible equivalent of
CALCulate<Chn>:MARKer<Mk>:FUNCtion:EXECute LPEak.
Channel number used to identify the active trace
Marker number in the range 1 to 10
Device-specific, no query
CALCulate<Chn>:MARKer<Mk>:SEARch:NEXT
[Deprecated command]
Selects a search mode for marker no. <Mk> and initiates a search for the next highest or lowest value
among the valid peaks. The marker must be created before using
CALCulate<Chn>:MARKer<Mk>[:STATe] ON.
Note: This command is the ZVR-compatible equivalent of