R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 946
setting in the segment, which is equivalent to the minimum sweep time possible.
Delay for each partial measurement in the segment. See
[SENSe<Ch>:]SEGMent<Seg>:SWEep:DWELl. In the setting
[SENSe<Ch>:]SEGMent<Seg>:INSert:SELect SWTime, this parameter is
replaced by <Time>.
0 s to 2.5E+003 s. [s] AUTO activates automatic sweep time setting in the segment,
which is equivalent to a meas. delay of 0 s.
Ignored parameter, should be set to the default value 0.
Resolution bandwidth in the segment. See
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution].
1 Hz to 1 MHz (5 (10) MHz with option R&S ZVA-K17, 10 (30) MHz with option R&S
ZVA-K7) [Hz].
Position of the local oscillator frequency LO relative to the RF frequency (Spur
Avoid). See [SENSe<Ch>:]SEGMent<Seg>:SBANd.
POSitive: LO > RF
NEGative: LO < RF
AUTO: The analyzer auto-selects the local oscillator frequency, depending on the
receiver (RF) frequency and the test port.
Selectivity of the IF filter. See
[SENSe<Ch>:]SEGMent<Seg>:BWIDth[:RESolution]:SELect.
Device-specific, with query (used to retrieve the channel settings for a particular
sweep segment).
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using default settings.
SEGM:DEF?
Query the channel settings for the new segment.
[SENSe<Ch>:]SEGMent<Seg>:DEFine:SELect SWTime | DWELl
Defines whether the sweep time of a new segment, i.e. numeric parameter no. 5 of the command
[SENSe<Ch>:]SEGMent<Seg>:DEFine, is entered as a segment sweep time or as a meas. delay.
Device-specific, command or query.