R&S
®
ZVA / R&S
®
ZVB / R&S
®
ZVT Command Reference
SCPI Command Reference
Operating Manual 1145.1084.12 – 30 954
Couple the power levels in all segments and reset the power in segment no. 1 to the
initial value.
[SENSe<Ch>:]SEGMent<Seg>:SBANd AUTO | POSitive | NEGative
Defines the position of the local oscillator frequency LO relative to the RF frequency (sideband, Spur
Avoid) in sweep segment no. <Seg>. At the same time, the command activates separate sideband
settings in all sweep segments ([SENSe<Ch>:]SEGMent<Seg>:SBANd:CONTrol ON).
The analyzer auto-selects the LO frequency, depending on the receiver (RF)
frequency and the test port.
Device-specific, command or query
See [SENSe<Ch>:]SEGMent:NAME
[SENSe<Ch>:]SEGMent<Seg>:SBANd:CONTrol <Boolean>
Selects common or independent LO sideband (Spur Avoid) settings for the sweep segments.
Sweep segment number. This suffix is ignored; the setting controls the whole
segmented sweep.
ON – Use independent sideband settings, to be defined via
[SENSe<Ch>:]SEGMent<Seg>:SBANd.
OFF – Reset the sideband in all sweep segments to the sideband for unsegmented
sweeps, defined via SENSe<Ch>:FREQuency:SBANd. ON will not restore the
previous values.
OFF. The parameter is automatically switched to ON when a sideband setting is
defined using [SENSe<Ch>:]SEGMent<Seg>:SBANd.
Device-specific, command or query
See [SENSe<Ch>:]SEGMent:NAME