Remote Control Commands
R&S
®
SMBV100A
707Operating Manual 1176.8016.02 ─ 17
Example:
BB:MCCW:TRIG:SOUR EXT
selects an external trigger
BB:MCCW:TRIG:DEL 200
sets a delay of 200 samples for the trigger.
Manual operation: See "Trigger Delay" on page 447
[:SOURce<hw>]:BB:MCCW:TRIGger[:EXTernal<ch>]:INHibit <Inhibit>
Specifies the number of samples by which a restart is to be inhibited following a trigger
event. This command applies only in the case of external triggering.
Parameters:
<Inhibit> integer
Range: 0 to 67108863
*RST: 0
Example:
BB:MCCW:TRIG:SOUR EXT
selects an external trigger
BB:MCCW:TRIG:INH 200
sets a restart inhibit for 200 samples following a trigger event.
Manual operation: See "Trigger Inhibit" on page 448
[:SOURce<hw>]:BB:MCCW[:TRIGger]:SEQuence <Sequence>
Selects the trigger mode.