2-8
IM DLM4038-02EN
Setting the Qualifications (Qualification)
Press the Qualification soft key to open a menu. The menu that appears varies depending on the 
specified trigger source.
 When the Trigger Source Is a Channel from CH1 to CH8 or LOGIC
  Example: When the Trigger Source Is CH1
 
Set the trigger slope for the trigger source signal. 
•   Select     ,     , or X when the window comparator is off. 
•   Select      ,      , or X when the window comparator is on. 
Set qualifications for signals other than the trigger source.
•   Select H, L, or X when the window comparator is off.
•   Select IN, OUT, or X when the window comparator is on.
/L16 option
When the Trigger Source Is EXT (External trigger signal)
  The same menu appears as that shown above for when the trigger source is a channel from CH1 
to CH8 or LOGIC. Because the trigger source is an external signal (EXT), you can specify all of the 
signal states for CH1 to CH8 and LOGIC as qualifications.
2.5  Triggering on Edge Conditions