9-6
IM DLM4038-02EN
9.2  Processing Statistics on Automatically 
Measured Values
This section explains the following settings (which are used when processing statistics on automatically 
measured waveform parameters).
•  Statistical processing mode
•  Normal statistical processing
•  Cyclic statistical processing
•  Statistical processing of history data
► “Statistics (Statistics)” in the Features Guide
MEASURE Statistics Menu
Press MEASURE, the Statistics soft key, and then the Mode soft key to display the following menu.
 
Configure statistical processing of history data. 
Configure cyclic statistical processing. 
Configure normal statistical processing.
Disable statistical processing.
Setting Normal Statistical Processing (Continuous)
Press the Continuous soft key to display the following menu.
 
Configure the trend display or the histogram display. 
Restart statistical processing.
Set whether statistical processing is restarted 
when the trigger level is changed.
Setting for Restarting Statistical Processing When the Trigger Level Is Changed
  Restart:  If the trigger level is changed during waveform acquisition, the statistical processing 
performed up to that point is discarded, the waveform Count is set to 1, and statistical 
processing restarts.
  Ignore:   If the trigger level is changed during waveform acquisition, waveform acquisition and 
statistical processing continue without statistical processing being reset.