Page #42 background image
2-10
IM DLM4038-02EN
Setting the Clock Source Pattern
Press the Pattern soft key to display a menu. The menu that is displayed varies depending on the
specified clock source.
When the Clock Source Is a Channel from CH1 to CH8 or LOGIC
Example: When the Clock Source Is CH1
Set the slope for the clock source signal.
Select or when the window comparator is off.
Select or when the window comparator is on.
Set the patterns for signals other than the clock
source.
Select H, L, or X when the window comparator
is off.
Select IN, OUT, or X when the window
comparator is on.
/L16 option
When the Clock Source Is Not Specified
The same menu appears as that shown above for when the clock source is a channel from CH1
to CH8 or LOGIC. Because no clock source is specified, you can specify all of the signal states
for CH1 to CH8 and LOGIC as state conditions.
2.6 Triggering on State Conditions