2-27
IM DLM4038-02EN
Setting Trigger Conditions (Condition Setup)
  Press the Condition Setup soft key to display the following screen.
  The DLM4000 triggers on the AND of the SOF, ID, frame type (Remote Frame or Data Frame), 
Data, and ACK conditions. Items whose check boxes are selected are used as trigger conditions.
•  When the Comparison Condition Is True or False
   
Set the ID bit pattern.
If you select Extend for the frame format, 29 
bits are displayed here.
Set the data pattern.
Set the comparison condition.
Set the comparison size.
Set the comparison start position.
Set the trigger source frame.
SOF (always selected)
Set the frame format.
Set the ID input format (Bin, Hex).
Set the data pattern input format.
Set the ACK slot state.
• WhentheComparisonConditionIsData=a;Data≠a;a≤Data;Data≤b;a≤Data≤b;or
Data < a, b < Data
   
Set data reference values a and b.
Set the comparison condition.
Set the byte order.
Set the ID bit pattern.
If you select Extend for the frame format, 29 
bits are displayed here.
SOF (always selected)
Set the frame format.
Set the ID input format (Bin, Hex).
Set the comparison size.
Set the comparison start position.
Set the trigger source frame.
Set the bit positions of the most 
significant bit (MSB) and the least 
significant bit (LSB) in the data that 
you will compare.
Set whether to use a signed (Sign) or 
unsigned (Unsign) data format.
2.11  Triggering on CAN FD Bus Signals (Option)