2-41
IM DL950-02EN
DLC Value and the Number of Transmission Data Bytes
When DLC = 0 to 8
DLC
Number of Data
Bytes
DLC
Number of Data
Bytes
DLC
Number of Data
Bytes
CAN CAN FD CAN CAN FD CAN CAN FD
0 0 0 1 1 1 2 2 2
3 3 3 4 4 4 5 5 5
6 6 6 7 7 7 8 8 8
When DLC ≥ 9
When the message type is CAN
DLC
Number of Data
Bytes
9 to 15 8
When the message type is CAN FD
DLC
Number of Data
Bytes
DLC
Number of Data
Bytes
DLC
Number of Data
Bytes
9 12 10 16 11 20
12 24 13 32 14 48
15 64
2.9 Configuring the Monitoring of CAN and CAN FD Bus Signals (/VCE option)