3/35 Sample Loader TS Guide 9140234B-June 1997
TROUBLESHOOTING
1.1 OVERVIEW
1 - 2
1
Search
Book TOC
Go Back
1.1 OVERVIEW
This chapter contains the followin
g
information
for Type E, F, & G Sample Loaders:
1.2 Error Messa
g
es
1.3 Isolation Procedures (IPs)
Troubleshooting Approach
Troubleshootin
g
is required whenever an
Observed Error or a Displayed Error occurs.
Sample Loader Displayed Errors can be
viewed on the Bulletin Line of the Data Station
screen. (The Bulletin Line is down about 3/4
way on the screen.)
When a malfunction occurs, follow these
recommended procedures:
1. Decontaminate the Sample Loader and the
Analyzer accordin
g
to General Biosafety
Procedure.
2. Use this Troubleshootin
g
Guide as follows:
a. If possible, be
g
in by verifyin
g
the
reported error. Perform the exact pro-
cedure that produced the error condi-
tion.
b. When an error occurs, either observed
or displayed, look up the error in Sec-
tion 1.2, Error Messa
es table.
c. Perform the Corrective Action.
d. If directed to an IP, flip pa
es while look-
in
on ed
e of sheets for the desired IP
Number.
e. Turn to pa
e 1 of x, and be
in at the
START box.
f. Follow the directions in the IP.
g
. If directed to perform a Verification Pro-
cedure (VP), flip pa
es while lookin
on
ed
e of sheets for the desired VP Num-
ber.
h. If a part is adjusted or replaced,
o to
Section 3.2, Interrelationship Matrix.
i. Find part in left column.
j. Perform VPs in order
iven (i.e. 1, 2, 3,
4, etc.)
3. Perform Total Call.
--- END ---