1997 20:45
CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
TROUBLESHOOTING
IP - 18 Alarm 4 (s) Sample Tube Not Sensed Properly, 4 of 4
1 - 73
1
Search
Book TOC
Go Back
IP - 18 Alarm 4 (s) Sample Tube Not Sensed Properly, 4 of 4
Replace LED
(W hite Beam Block PL-6.1), or
(Black Beam Block PL-2.5)
(See Interrelationship Matrix - Pg. 3-4)
Voltage
Increase to
> 2.0V With
Light
?
Replace Detector
(W hite Beam Block PL-2.5), or
(Black Beam Block PL-6.1)
(See Interrelationship Matrix - Pg. 3-4)
No
Yes