3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURES
VP - 3.17 X-Axis Solenoid to Rack Guide Adjustment
3 - 57
3
Search
Book TOC
Go Back
VP - 3.17 X-Axis Solenoid to Rack Guide Adjustment
Purpose:
Perform this procedure to ali
g
n X-Axis Solenoids with Rack Guides.
1. Perform Check procedure “VP - 3.16 X-Axis Solenoid to Rack Guide Check” (Pa
g
e 3-55).
CAUTION:Do not loosen Rack Guides to perform this adjustment.
2. Adjust Home Sensor for proper ali
g
nment of Solenoids to Ri
g
ht Rack Guide (Fi
g
ure 3-25,
Pa
g
e-3-56 and Fi
g
ure 3-26, Pa
g
e-3-58).
a. Loosen Ri
g
ht (Home) Sensor screws.
b. Adjust Sensor bracket until X-Axis Solenoids ali
g
n with Rack Guide.
c. Put Sample Loader in Manual Mode
1.
Press and Hold
MAN
and
START
.
2.
Turn Power
ON
, hold keys down until BEEPING stops.
3.
Remove racks.
4.
Ensure X-Axis Solenoids are ali
g
ned with Ri
g
ht Rack Guide. Check with a strai
g
ht ed
g
e.
d. Check Ri
g
ht Rack Guide by movin
g
X-Axis Solenoids (
X
) 1/2” (1cm) from Ri
g
ht Stop (hi
g
h
motor speed). Press (
X
) to move Solenoid to Ri
g
ht Stop (Home) in small steps (slow motor
speed) to prevent Sensor overshoot.
e. Ensure neither Solenoid extends beyond the Rack Guide. Check with a strai
g
ht ed
g
e (Fi
g
ure
3-25, Pa
g
e-3-56).