28 Oct
1997 21:57
CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURES
VP - 3.4 Rack Clearance Check
3 - 26
3
Search
Book TOC
Go Back
VP - 3.4 Rack Clearance Check
Purpose:
Perform this procedure to verify clearance between Rack and Beam Block, Fi
ure 3-9.
1. Check Clearance of Rack at Beam Block (Fi
ure 3-9).
2. If Check is OK,
o to “VP - 3.6 X-Axis Solenoid to Guide Rod Check” (Pa
e 3-31).
3. If Check Fails,
o to “VP - 3.5 Rack Clearance Adjustment” (Pa
e 3-28).
Special Tools
DESCRIPTION QTY PN
Feeler Gau
e Set 1 1-43627-01