3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURES
VP - 6.1 Mixing Head Position Check
3 - 77
3
Search
Book TOC
Go Back
VP - 6.1 Mixing Head Position Check
Purpose:
Perform this procedure to verify ali
g
nment of Mixin
g
Assembly.
1. Perform Mixin
g
Head Vertical Check as follows;
a. Remove Racks.
b. Access rear of Loader.
c. Ensure Ball Slide mountin
g
screws are ti
g
ht (Fi
g
ure 3-36).
Special Tools
DESCRIPTION QTY PN
Mixin
g
Head
Adjustment Tool
1 9420082
Rack Guide
Ali
g
nment Tool
1 9420084