CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURES
VP - 6.3 Mixing Home Sensor Check/Adjustment
3 - 89
3
Search
Book TOC
Go Back
VP - 6.3 Mixing Home Sensor Check/Adjustment
Purpose:
Perform this procedure to verify and adjust Mixin
Head Assembly Home Position.
All Loaders
1. Clear Racks from under Mixin
Paddle.
2. Put Sample Loader in Manual Mode.
a. Press and Hold
MAN
and
START
.
b. Turn Power
ON
, hold keys down until BEEPING stops.
3. Press (
↓ Z ↑
) keys to move the Mixin
Paddle all the way down then all the way up and check
measurement (Fi
ure 3-44).
NOTE:Measure from knurled edge of Mix Paddle not rounded tip.
4. If check fails, adjust Home Sensor (Fi
ure 3-44).
NOTE:If check continues to fail, ensure Shims (2) are present below Mixing Assembly. See TSB
97-030.
5. Go to “VP - 6.4 Mixin
Contact Sensor Check/Adjustment” (Pa
e 3-91).
Special Tools
DESCRIPTION QTY PN
None