3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURES
VP - 9.2 Vent Wash Block Check/Adjustment- (F Loader Only)
3
Search
Book TOC
Go Back
VP - 9.2 Vent Wash Block Check/Adjustment- (F Loader Only)
Purpose:
Perform this procedure to verify and adjust Vent Needle Ali
nment, F Loader Only, Fi
ure
3-70.
1. Perform Steps 1, 2, and 3 of “VP - 9.1 Sample Wash Block Check/Adjustment” (Pa
e 3-125).
2. Ensure proper Needle is installed and Needle Port is centered on Wash Port (Fi
ure 3-69).
Special Tools
DESCRIPTION QTY PN
Needle Ali
nment
Tool
1 9420081
Rack Guide
Ali
nment Tool
1 9420084