28 Oct
1997 21:57
CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURES
VP - 3.2 Platen Wall Check
3 - 20
3
Search
Book TOC
Go Back
VP - 3.2 Platen Wall Check
Purpose:
Perform this procedure to verify Platen Walls are square to one another and Rack Guides
and Beam Block are ali
ned.
1. Check fit of Rack Guide Ali
nment Tool in Positions 1,2,3,4.
• If Check is OK,
o to “VP - 3.4 Rack Clearance Check” (Pa
e 3-26).
• If Check Fails,
o to “VP - 3.3 Platen Wall Adjustment” (Pa
e 3-22).
Special Tools
DESCRIPTION QTY PN
Rack Guide
Ali
nment Tool
1 9420084